 Abstract

In this paper, a practical implementation to reduce leakage flux of a high-frequency inverter based non-contact type power transformer composed of EE-shape ferrite cores is presented for key technology of the next generation medical use X-ray CT scanner system. Design consideration for the unique structure of the non-contact power transformer with 900mm in diameter is also introduced. The complete non-contact transformer is actually arranged by several blocks of the magnetic circuit assembled by using 10 small EE shape cores with 120mm in length. It is experimentally and analytically discussed from a reduced leakage flux viewpoint related to its inductively coupling coefficient. A practical method to lower the leakage flux is described based on effective Copper-Sheet-Treatment placed on EE shape ferrite cores of magnetic circuit.


 Statistics
Show / Hide Statistics

Cumulative Counts from September 30th, 2019
Multiple requests among the same browser session are counted as one view. If you mouse over a chart, the values of data points will be shown.



Cite this article

[IEEE Style]

L. Gamage, M. Isitobi, M. Nakaoka, "Performance Evaluation a Novel Prototype of High Frequency Non-Contract Power Transformer," Journal of Power Electronics, vol. 4, no. 1, pp. 12-17, 2004. DOI: 10.6113/JPE.2004.4.1.12.

[ACM Style]

Lanknath Gamage, Manabu Isitobi, and Mutsuo Nakaoka. 2004. Performance Evaluation a Novel Prototype of High Frequency Non-Contract Power Transformer. Journal of Power Electronics, 4, 1, (2004), 12-17. DOI: 10.6113/JPE.2004.4.1.12.